제주대학교 Repository

투과 반사 통합형 디지털 홀로그래피 현미경에 관한 연구

Metadata Downloads
Alternative Title
Study of Digital Holographic Microscope Using a Combination of Transmission-Reflection
Abstract
Since nano technology was introduced recently, it has been used in studies on material characteristics of non-organism such as carbon nanotube, single electron transistor and semi-conductor. However lately nano technology has been applied to biology for understanding physical and chemical phenomenon of a molecular. It is expected to raise the revolutionary change throughout bio technology such as biology and medicine. But, the past nano technology has been restricted for applying to biology and medicine which must observe a living body in real time.

Digital holographic technology has been developed for the last three decades to apply to In-Vivo inspection, encrypting technology, 3D vision, and holographic microscope. We call it 'digital holographic technology' that reconstructs 3D image of hologram recorded in CCD(Charged Coupled Device) with using numerical diffraction calculation. This technology was proposed by Goodman and Laurence, and realized by Kronrod et al. The biggest difference in classic holography and digital holography is the hologram recorder and 3D image reconstruction method. A recorder can abbreviate chemical hologram process, and record the hologram data in real time with using digital image recorder such as CCD instead of holography photo-plate. Also digital holography can get spatial benefit, and moreover, save, copy and edit the measured results. And unlike classic holography that reconstructs 3D image by optical method, digital holography reconstructs 3D image by computer. Digital holographic technology has more interest in large number of applications for a high resolution real-time 3D observation when the technology of computer & digitial recorder has developed remarkably.
However, we have some difficulty in embodying the image with good signal to noise ratio because DC term including zero-order suppression and twin image work as noise when the image is reconstructed numerically with using hologram from CCD. Holography is classified with an in-line hologram and an off-axis hologram by the incidence angle of reference beam. Gabor developed in-line hologram for the first time. Its incidence angle of reference beam to objective beam is 0°.
In the in-line hologram, it is difficult to provide a clear image because real image, virtual image, and DC term can be superposed. These problems were solved by off-axis hologram designed by Leith and Upatnieks. We call it when reference beam enters with angle, and real image, virtual image, and DC term are separated spatially. However, a real image in off-axis is reconstructed within only 1/4 of reconstruction zone. For that reason, it has a bad effect on FOV(Field Of View) of digital holography because resolution of digital holography recorder is less than 50 lines/mm . Virtual image and DC term must be removed with using in-line hologram to get a high resolution image.
Also, in case of measurement using the light, refractive index is an important factor. The 3D depth information is distorted as much as the difference in refractive index because each organizer in a living body provides the various refractive index. Furthermore, refractive index can give the information to convert the density of materials reversely. This is useful information in test of sorts, densities of a living body, and defect of materials. Only the Ellipsometry can measure object's structure and refractive index simultaneously. However, Ellipsometry can provide a fragmentary information of film thickness, but composes a 3D image. Also its measuring time is very long because it is a scan type. Until now, there is no report about any study on measuring a real 3D structure and a refractive index simultaneously.

This study was performed to construct an in-line digital holographic microscope to embody the highest resolution of CCD, and to research the method of removing integrated virtual image and DC term effectively. And, in this study I measured 3D structure and refractive index simultaneously by constructing digital holographic microscope using a combination of transmission-reflection.
At first, for the case of removing virtual image, I researched the virtual image removal method using geometric arrangement and the virtual image removal method using a zero padding method. The virtual image removal method using geometric arrangement is to use the phenomenon that virtual image size is minimized with considering the distance from lens to CCD, and the distance from CCD to a reconstructed plane. This method uses specific property in holography using a lens. However, in case of removing real image with a geometric method, there is some constraint in the measurement and reconstruction process. We studied the virtual image removal method using a zero padding method that was studied to complement this problem. The virtual image removal method using a zero padding method uses also specific property in holography using a lens. In an in-line digital holographic microscope, because the interference information is recorded in a hologram with some local limitation, the reconstructed image also is locally limited. That is to say, if we reconstruct each of hologram after dividing hologram into 4 domains, then we can get each of reconstruction image whose virtual image and real image are separated. Futhermore, if we combine 4 domains after selecting only the real images, then we can get a 3D image without virtual image. The virtual image removal method using a zero padding method has no dependence on hardware, and also exerts immaterial influence on reconstruction speed.
In case of removing DC term, because its intensity is very strong, much more than that of virtual image, DC term must be removed to embody the image with higher signal to noise ratio. In this study, it researched zero-order suppression method using scanning, zero-order suppression method using object beam & reference beam, and zero-order suppression method using the average removal method to remove DC term. Zero-order suppression method using scanning is to construct a hologram selectively by moving a subject of zero-order suppression, and remove zero-order beam.
When using this method, it is inconvenient more than other methods because it needs experimentally repetitive hologram saving and repetitive calculation, but it can remove DC term effectively. Zero-order suppression method using object beam and Reference Beam is to eliminate the intensity of reference beam and object beam from the hologram intensity. This method can provide a hologram without the DC term removed by eliminating zero-order beam and object beam. Because this method doesn't use a scanning method, it gives repetitive work inconveniency. But an object beam and a reference beam must be measured whenever we measure a subject. And refractive index of zero-order beam is significant as compared to scanning. Zero-order suppression using the average removal method is used widely because it can remove DC term with measuring a hologram once the reconstruction speed is fast, and the removal efficiency of DC term is relatively higher. The average removal method is to calculate the average of hologram intensity and then to extract the average for removing DC term. However, if we lower the average range(number of pixel to calculate the average), then DC term elimination rate will be improved. But at the same time, the high frequency region will be removed too. I researched the variable average removal method. This method is applied with the average range within a low frequency region and a high frequency region of hologram by grade. In consequence, we have minimized the loss of real image with maintenance DC term elimination rate.
Furthermore, I developed a digital holographic microscope using a combination of transmission-reflection by combining transmission holographic microscope that is influenced by refractive index and reflection holographic microscope that is not influenced by refractive index, and then measured the 3D images and refractive index simultaneously.
Author(s)
趙亨峻
Issued Date
2008
Awarded Date
2008. 8
Type
Dissertation
URI
http://dcoll.jejunu.ac.kr/jsp/common/DcLoOrgPer.jsp?sItemId=000000004480
Alternative Author(s)
Cho, Hyung-Jun
Department
대학원 물리학과
Advisor
劉榮勳
Table Of Contents
제 1 장 서 론 1

제 2 장 이론적 배경 5
1. 홀로그래피 5
1.1 홀로그램 기록 및 재생 6
2. 디지털 홀로그래피 현미경 9
2.1 디지털 홀로그램 기록 10
2.2 디지털 홀로그램 재생 13
2.3 위상 펼침 16
3. In-Line 디지털 홀로그래피 20
3.1 허상 제거 20
3.1.1 기하학적 배치를 이용한 허상 제거 21
3.1.2 Zero Padding을 이용한 허상 제거 21
3.2 영차 회절 광 제거 25
3.2.1 평균 제거 방법을 이용한 영차 회절 광 제거 26
4. 투과 반사 통합형 디지털 홀로그래피 현미경을 이용한 굴절률 측정 27

제 3 장 실험 및 방법 33
1. 일반적인 투과형과 반사형 In-Line 디지털 홀로그래피 현미경 33
2. 투과 반사 통합형 디지털 홀로그래피 현미경 35

제 4 장 결과 및 고찰 40
1. In-Line 디지털 홀로그래피에서 허상 제거 40
1.1 기하학적 배치를 이용한 허상 제거 40
1.2 Zero Padding을 이용한 허상 제거 43
2. In-Line 디지털 홀로그래피에서 영차 회절 광 제거 51
2.1 주사(Scan) 방법을 이용한 영차 회절 광 제거 51
2.2 물체 광과 기준 광을 이용한 영차 회절 광 제거 57
2.3 변형된 평균 제거 방법을 이용한 영차 회절 광 제거 62
3. 프레넬 Zone Plate 효과 제거 90
4. 투과 반사 통합형 홀로그래피 현미경을 이용한 굴절률 측정 98
4.1 굴절률 측정의 전산기시늉 99
4.2 PCX 렌즈의 굴절률 측정 107
4.3 Cr-less 격자무늬 시료의 굴절률 측정 114

제 5 장 결 론 120

Abstract 123

참고문헌 128
Degree
Doctor
Publisher
제주대학교 대학원
Citation
趙亨峻. (2008). 투과 반사 통합형 디지털 홀로그래피 현미경에 관한 연구
Appears in Collections:
Faculty of Applied Energy System > Physics
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.