Abstract
Twinkling Artifact (TA) is an artifact which appears with random color
changes in red and blue, even if there is no turbulence or a similar movement
in the tissue. TA is often observed in renal stone, calcification, chronic
pancreatitis accompanied with calcification, papilloma thyroid carcinoma, and
calcinosis. The underlying mechanisms of TA have not been clearly
understood, yet. In this study, the sources that is expected to cause TA was
classified into target characteristics and machine settings, their effects on
TA were experimentally tested using phantoms. In particular, this study
proposed a new technique for enhancing TA contrast and evaluated its clinical
potential. The present study is divided into three areas: (1) the effect of
target characteristics, (2) the effect of color Doppler setting parameters, and
(3) the TA contrast enhanced by mechanical impulse.
First, this study examined the influence of various scatter characteristics of
target on TA using tissue mimicking (TMM) phantoms under the same color
Doppler settings. It was found that TA occurred more as the concentration
and size scatterers were increased.
Second, the effects of color Doppler settings on TA were assessed using a
circular contrast targets, for the following eight setting parameters of PRF,
frequency, WMF, ensemble, line density, smoothing, gain and CWP. TA was
observed to be maximized when PRF, frequency, WMF and ensemble were
low, line density, smoothing and CWP were high, and gain was mid.
Third, this study found that TA increased rapidly at the moment when
mechanical impulse was applied to the target. The increased TA induced by
the impulse persisted longer when the impulse was stronger and the target
contrast was high. This finding was verified to a patient with tiny renal stones
which are hard to be seen on conventional ultrasonic images.
In conclusion, the present study shows that TA is closely related to target
characteristics and was affected by the color Doppler settings. The study
suggested the use of a mechanical impulse to enhance TA and if such a
'Contrast Enhanced TA Mode' was implemented in an ultrasonic scanner. TA
is expected to confirm microcalcification and fibrocystic lesion (i.e. early breast cancer, chronic infection, and tiny stone) which cannot be clearly
identified with conventional ultrasonic images.
Author(s)
양정화
Issued Date
2014
Awarded Date
2014. 8
Type
Dissertation
URI
http://dcoll.jejunu.ac.kr/jsp/common/DcLoOrgPer.jsp?sItemId=000000006904
Alternative Author(s)
Yang, Jeong-hwa
Department
대학원 의공학협동과정
Table Of Contents
Ⅰ. 서 론
1.1 배경 1
1.2 연구동기 및 필요성 6
1.3 연구목적 9
1.4 논문의 구성 9
Ⅱ. TA의 기전 및 의학적 유용성
2.1 서론 11
2.2 TA 발생기전 12
2.2.1 타겟 특성과의 상관성 13
2.2.2 컬러 도플러 설정이 미치는 효과 19
2.3 임상적 유용성
2.3.1 결석 (calculus) 20
2.3.2 석회화 (calcification) 23
2.3.3 석회증 (calcinosis) 29
2.3.4 다양한 병변 29
2.3.5 고에코성 이물질 (echogenic foreign bodies) 33
2.4 고찰 35
2.5 결론 38
Ⅲ. 타겟의 특성에 의한 효과
3.1 서론 39
3.2 실험
3.2.1 팬텀 42
3.2.2 초음파 영상 획득 45
3.2.3 TA의 정량적 평가 45
3.3 결과 47
3.4 고찰 50
3.5 결론 59
Ⅳ. 컬러 도플러 설정변수가 TA에 미치는 효과
4.1 서론 60
4.2 컬러 매핑과 컬러바 61
4.3 컬러 도플러 설정변수 64
4.4 실험 71
4.4.1 타겟 팬텀과 초음파 영상 획득 73
4.4.2 컬러 도플러 설정변수 73
4.5 결과
4.5.1 타겟 대조도의 효과 74
4.5.2 타겟 도플러 설정변수의 영향 76
4.6 고찰 89
4.7 결론 99
Ⅴ. 충격 자극에 의한TA 대조도 개선
5.1 서론 101
5.2 연구방법 및 실험
5.2.1 실험장치 104
5.2.2 기계적 충격 자극 106
5.2.3 TA의 정량적 평가 지수 108
5.2.4 실험 방법 110
5.3 결과 110
5.3.1 충격파 세기에 따른 TA 발생 특성 113
5.3.2 타겟의 대조도에 따른 TA 발생 특성 113
5.3.3 타겟 내 TA 발생량의 시간에 따른 변화 114
5.3.4 타겟의 TA 영상의 대조도 117
5.4 고찰 124
5.5 결론 134
Ⅵ. 결론
6.1 주요 결과 135
6.1.1 타겟의 특성이 TA에 미치는 효과 135
6.1.2 컬러 도플러 설정변수가 TA에 미치는 효과 136
6.1.3 충격 자극에 의한 TA 대조도 개선 137
6.2 추후연구 138
6.3 결어 140
참 고 문 헌 141
요 약 152
Degree